SF - Immunization Devices
Download file
Product Details
Product name
LMPro T051
IMD-PQS Code
E006-100
Status
Prequalified
Date of acceptance
25 Nov 2024
Product description
User-programmable temperature data logger
Manufacturer
G-Tek Corporation Private Limited
Manufacturer Reference
LMPro T051
Country of Manufacture
India
Address
Gunaji House”, Plot No. 25/1, Besides Status Bungalow, Padra Road,
Vadodara
Gujarat
India
Vadodara
Gujarat
India
Telephone
+91 98 245 22295
Email
Website address
Valid until
31 May 2026
Specifications
Product specifications - Main
Mode of operation
Electronic
Minimum logging intervals (mins)
0.00
External dimensions height (mm)
95.50
External dimensions length (mm)
13.80
External dimensions width (mm)
45.00
Material
Polycarbonate plastic; non- breakable, non-corrodible
Minimum working temperature (°C)
-30.00
Maximum working temperature (°C)
60.00
Mechanical and electronic devices only
Power source
Other
Power Source (Other)
Lithium Coin Cell Battery
Battery type
User Replaceable 3.0 V 225 mAH; CR2032 Panasonic
Battery shelf life (year)
1
Activated battery life
12 months
Alarm type
Visual
Memory capacity (readings)
50000.00
Programmability
User Programmed
User interface
Keyboard interface
Price and shipping
Shipping weight per carton(kg)
2.5
Shipping volume (m3)
0.018
Units per package
25.00
Minimum order
1.00
Incoterms
EXW
Price 1
For 1-10 units US$49 per unit
Price 2
For 100-250 units US$34 per unit
Price 3
For more than 250 units :Consult Factory
Year of base price
2024
Quality standard
Quality Standard
Other
Quality standard (Other)
ISO 9001:2015
Verification
Verification Laboratory
0013X00004Ml8OlQAJ
Verification Report Reference
LLTPL/TL/2024-02/01
Document Details

LMPro T051 User Programmable Temperature Data logger is an Internal sensor data logger meeting the requirements of WHO PQS E006/TR05.1. It can store up to 50000 data records. User can fully configure parameters like – Start delay, alarm high/low with corresponding alarm delay, alarm event type- single/cumulative, temperature unit, UTC time zone, store interval, Batch type- Stop on full or memory rollover etc. User can also set the information like, username and job description. It displays OK symbol “Tick” if the temperature has remained within the recommended limits and NOK symbol “Cross” if there are violations for the set alarm limits. User can also download data from multiple data loggers in single file and generate report in pdf format or export data as a csv file.