SF - Immunization Devices
Download file
Product Details
Product name
VaxAlertTM USB
IMD-PQS Code
E006-029
Status
Prequalified
Date of acceptance
18 Dec 2014
Appliance type
E006
Product description
40 day electronic shipping indicator
Manufacturer
Sensitech Inc.
Manufacturer Reference
VaxAlertTM USB
Country of Manufacture
China
Address
A Business Unit of Carrier Corporation, Suite 258X, 800 Cummings Center. Beverly, MA 01915 United States
United States of America
United States of America
Telephone
+1 978 - 927- 7033
Email
Website address
Valid until
31 May 2026
Specifications
Product specifications - Main
Mode of operation
Electronic
Minimum logging intervals (mins)
2.00
Maximum recording time (days)
20.00
Sensor type(s)
Electronic digital
Maximum sensor lead length (metres)
0.00
Maximum number of sensor channels
1.00
External dimensions height (mm)
19.60
External dimensions length (mm)
101.00
External dimensions width (mm)
56.00
Activation mechanism
Start/Stop button
Material
ABS
Fixing mechanism
None
CE marking
Yes
Minimum working temperature (°C)
-30.00
Maximum working temperature (°C)
55.00
Mechanical and electronic devices only
Accuracy (+/- °C)
0.50
Power source
Non-replaceable battery
Battery type
3.6V Lithium
Battery shelf life (year)
24 mths
Activated battery life
12 mths
Alarm type
Visual
Memory capacity (readings)
16000.00
Programmability
Factory Programmed
User interface
LCD and PC downloadable
Display medium
LCD
Recording medium
Electronic
Price and shipping
Weight per Package (Kg)
0.0709
Units per package
20.00
Weight per Unit (Kg)
1.0400
Minimum order
20.00
Incoterms
EXW
Price 1
21.00€ per unit
Year of base price
2014
Quality standard
Quality Standard
Other
Quality standard (Other)
ISO 9001:2015
Specification Reference
Applicable PQS specification: E006/TR07.3
Applicable PQS VP(s): E006/TR07-VP.3
Applicable PQS VP(s): E006/TR07-VP.3
Verification
Verification Report Reference
ACS-E14535
Product Sites
Sensitech Inc.
A Business Unit of Carrier Corporation, Suite 258X, 800 Cummings Center. Beverly, MA 01915 United States
United States of America
Document Details

The indicators are factory configured with WHO established alarm profiles in support of UNICEF vaccine procurement and transport requirements. These cost-effective, single-use indicators monitor up to three time-temperature alarms, which can be tailored to specific cold chain transport monitoring requirements beyond WHO/UNICEF vaccine shipments. Its main features are: Automatic generation of Adobe® PDF cold chain data report; Integrated USB connector; Two types of alarms: (1) Single event alarms are triggered if temperature exceeds threshold for a contiguous time period, (2) Cumulative time event alarms are triggered if temperature exceeds threshold for a total sum of time. User-friendly three button interface and intuitive display enables immediate accept/reject decision making and detailed excursion history information. Alarm “time stamp” records elapsed time at point of alarm events, allowing supply chain partners to analyze variability and institute corrective actions.
Accuracy: ±0.5°C between -5°C and +25°C
Maximum Logging Intervals: Optional up to 2 hours