SF - Immunization Devices
Download file
Product Details
Product name
GHR 120 AC
IMD-PQS Code
E003-149
Status
Prequalified
Date of acceptance
04 Mar 2026
Appliance type
E003
Product description
Icelined refrigerator
Manufacturer
Godrej & Boyce Manufacturing Company Limited
Manufacturer Reference
GHR 120 AC
Country of Manufacture
India
Address
Plant 11, Pirojshanagar, Vikhroli
Mumbai
Maharashtra
India
Mumbai
Maharashtra
India
Telephone
+91 22 67966346
Email
Website address
Valid until
31 May 2027
Specifications
Product specifications - Main
Climate zone
Hot
Minimum rated ambient temperature (°C)
10.00
Hold over time (hours:mn)
79 hours
Appliance tested at (°C)
43.00
Performance at" (°C)
43.00
Ambient temperature during testing (°C)
43.00
Fuel and refrigeration cycle
Compression
Energy source
AC Power
Lining
PCM
Energy Consumption Intermittent Power (kWh/24h)
1.2200
Energy Consumption Cool Down (kWh/24h)
1.3800
Energy Consumption Continuous Power (kWh/24h)
0.5500
Product specifications - Additional Refrigerator
Refrigerator vaccine storage capacity(L)
118.000
Refrigerator's gross volume (Litres)
142.000
Autonomy (hrs:mn)
79 hours
Refrigerant type
R290
Solar radiation ref period (kWh/m2/day)
0.00
Comments
APPROVED EMS LEVEL 3
Price and shipping
Shipping volume (m3)
1.01
Shipping weight (kg)
145.00
Minimum order
1.00
Units per package
1.00
Incoterms
EXW
Price 1
1835
Price 2
1800
Price 3
1765
Year of base price
2026
Quality standard
Quality Standard
ISO 9001: 2000
Verification
Verification Report Reference
4791893252-02-OTHER-S3
Product Sites
Godrej & Boyce Manufacturing Company Limited
Plant 11, Pirojshanagar, Vikhroli
Mumbai,
Maharashtra
400 079
India
Document Details